← Back to context

Comment by moralestapia

8 hours ago

This would be cool if only it made sense.

To use an analogy with some metaphors: The sensor is like a sealed room with a screen window that only lets in oxygen. To get a reading, every molecule that enters is smashed to create a tiny spark of electricity. However, because the oxygen is destroyed to create that spark, it creates a suction effect, causing more oxygen to rush into the room to fill the void. This creates a major flaw: if gunk builds up on the screen, it slows down the flow of incoming oxygen. The sensor, which only counts sparks per second, is tricked into thinking the oxygen level outside is low, when really the window is just dirty.

By adding a third electrode to replace the oxygen every time one is smashed, you maintain a perfect balance and eliminate that suction. Because the room stays full, the sensor no longer relies on the speed of the oxygen rushing in; it simply measures the steady state of the oxygen already there. Even if gunk gets on the window, the sensor won't be starved of a reading. It might take a few extra seconds for the levels to settle, but the final number will be 100% accurate because the sensor is no longer emptying its own room to get a count.

  • I still don't get it. The outside is dirty, right? He said in his post "You dip this probe into beer, sewage, or canned food a-stewing". So when you say "when really the window is just dirty" I don't get it - yes it will always be, because that's what it is placed in, no?

    • A dirty window only ruins the reading if you are measuring the speed of the oxygen passing through it. The three electrode design stopped measuring speed and started measuring balance. Unless the gunk is a total airtight seal (which is rare on the scale of an oxygen molecule), the sensor will eventually reach the right answer, whereas the old version would fail.

      4 replies →

I agree. It's not clear how adding a sensor "so that it adds back an oxygen molecule" works. shrug

  • I think this was primarily about speeding up the measurement time. With just two electrodes you had to wait for the device to achieve equilibrium with the material being measured. If the concentration of oxygen on the probe side of the barrier was higher or lower than the material side you would get false measurements, particularly in low oxygen scenarios because you have oxygem trapped in the probe.

    By keeping the state of oxygen inside the probe constant and replacing consumed molecules you now can measure almost instantly.

    • Yes but how do you do that? that magical third electrode sounds harder to make than the original problem.

      Edit: I think I get it now, it's a chemical reaction. By applying a voltage with some polarity to the 3rd electrode you can run the reaction in reverse. Still very hard to achieve because you have to make sure the reactions happen at the same rate with the same efficiency, which is far from trivial. This must be a very high end sensor for all this effort to make sense.

      1 reply →

  • Because then it doesn't alter the side of the membrane where it does the reading (plus one minus one equals zero). That makes the measurement more accurate.

    • Specifically, if you assume a partial pressure of Oxygen and of all other gases on the electrode-side of the diffusion membrane, then you'll only see a certain number of "ionization events" per time, and you're limited in how much electrical signal you get by how fast oxygen can diffuse across the membrane. This is likely driven by maintenance of a partial pressure within the membrane. However if you re-ionize the oxygen that you deionized, then the partial pressure is much closer to equilibrium, and therefore the partial pressures are only dependent on the amount of oxygen outside of the membrane instead of being dependent on both the ionization rate and the recovery rate through the membrane. It probably makes the calculation a lot faster and more closely dependent on the environmental presence of oxygen which is what you want.

The current is measuring the rate of the reaction. With the two-terminal design the rate of the reaction is proportional to the rate of diffusion of the oxygen into the area where the reaction is taking place, which is related to the oxygen concentration around it but also can be affect by other things. With the third electrode, the current is proportional to the concentration of oxygen in the area around the sensor directly, which will equalise with its surroundings much more consistently than the rate of diffusion.

(A quick google brings up this document which describes the principle. No idea if this is the company in the story: https://semeatech.com/uploads/Tech_Docs/AN%20161205.pdf )

This way you're measuring change in oxygen concentration. As more oxygen comes into the compartment in order to equalize with the outside you consume and at the same time produce more oxygen. You measure the change in rate of oxygen consumption/production. It is always consuming/producing oxygen but the rate changes with the concentration.

At least that's what I assume.

  • I think of it differently.

    Before, you measured diffusion rate of oxygen and inferred oxygen concentration from that (the concentration outside the chamber is always greater than the concentration inside). Dirty membranes etc all changed the rate of diffusion, which caused issues.

    After you measure oxygen concentration directly (the concentration inside and outside the chamber are always the same).

Trust me, if we all understood Richard Feynman the first time he said something, the world would be a very different place.